Home >> Products>Products by Category>Colour & Appearance Testing Instruments>Reflectance Spectrophotometer IRS-1301
Catalogue number : 1501505-04
Reflectance Spectrophotometer IRS-1301 using a 2° or 10° viewing angle, 2° is used for viewing angles of 1 to 4; whereas 10° is used for viewing angles of more than 4. Reflectance spectrophotometer used d/8° geometry of sphere measurement diffused illumination integrating systems illuminate an object equally from all directions or angles and detects the reflected light at 8°.
Reflectance Spectrophotometer IRS-1301 using a 2° or 10° viewing angle, 2° is used for viewing angles of 1 to 4; whereas 10° is used for viewing angles of more than 4. Reflectance spectrophotometer used d/8° geometry of sphere measurement diffused illumination integrating systems illuminate an object equally from all directions or angles and detects the reflected light at 8°.
The applications including plastic, electronic, painting, coating, ink, textile, garment, printing and dyeing, food, medical, cosmetic, industries, research and development, and laboratories. Also, measure the reflectance spectrum and another color index precisely and control product quality management correctly. ISO and CIE recommended and equipped with a combined LED light sources with flat grating spectrophotometric mode. The IRS-1000 is used to measure specular component excluded (SCE) or specular component included (SCI) mode.
Accessories
Optional Accessories
Model |
IRS-1301 |
Wavelength |
400 to 700 nm |
Geometry |
d/8 (diffused illumination, 8°viewing angle) |
Illuminant |
D65,A, C, D50,D55,D75,F1,F2(CWF),F3,F4,F5,F6,F7(DLF),F8,F9,F10(TPL5)F11(TL84), F12(TL83/U30) |
Spectrophotometric Mode |
- |
Display |
3.5-inch TFT color LCD, Capacitive Touch Screen |
Detector |
Silicon photodiode array (double row 40 groups) |
Repeatability |
Spectral Reflectance: standard deviation within 0.1 %(400~700nm: within 0.2%) Colorimetric Value: Standard deviation within ΔE*ab 0.04 (Measurement conditions: white calibration plate measured 30 times at 5 seconds intervals after white calibration |
Wavelength Interval |
10nm |
Display Data |
Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, PASS/FAIL Result, Color Offset |
Operating Temperature |
- |
Storage Temperature |
- |
Source |
Combined LED sources |
Index |
WI(ASTM E313,CIE/ISO,AATCC), YI(ASTM, D1925,ASTM 313),TI(ASTM E313,CIE/ISO), Metamerism Index (Mt), Strength, Color Stain, Color Fastness |
Color Difference |
ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*cmc(l:c),CIE2000ΔE*00,ΔE(h) |
Color Space |
- |
Reflectance Range |
0-200% |
Observer |
2°/10° |
Battery |
Li-ion battery. 2800 times within 8 hours. |
Humidity |
0 to 85%R |
Interface |
USB/RS-232 |
Storage |
1000 Standards, 15000 Samples |
Dimensions |
90x77x230mm |
Weight |
Approx 600g |
Catalogue No. |
1501505-04 |