Home >> Products>Products by Category>Microplate Equipment>Microplate Reader IMR-4000(110V/50Hz)
Catalogue number : 420250501
Microplate Reader IMR-4000is asingle mode microplate readers are used for detection technology with 8-channel manifolds as standard accessories for use with standard 96 wells plates.
Microplate Reader IMR-4000is asingle mode microplate readers are used for detection technology with 8-channel manifolds as standard accessories for use with standard 96 wells plates.
Covers the absorbance range 0 - 4.000Abs with bandwidth <8nm and measured quickly, which is significantly faster than other current microplate reader methods. End Point, fixed time, Kinetic make the IMR-4000 the best platform for any clinical, biotechnology research and pharmaceutical laboratory applications. Can stored up to 100 programs and 100,000 test records in a single unit.Maintain all important measuring parameters including ABS, Cut-Off, Single Standard, Curve, Multi- Percent Log, Linear, Exponential, Power, 4PL Regression. A multi-mode reader has several benefits for laboratories applications ranging from ELISAs to nucleic acid and protein quantitation to cell imaging. Together with multiple microplate services and detection modes into a single, more versatile unit may be an ideal choice especially if you have limited area and performed 12 assays in a single plate.
| Model | IMR-4000 |
| Absorbance Range | 0 - 4.000Abs |
| Optical Path | 8-channel |
| Accuracy | ±0.5% or ±0.005Abs |
| Resolution | 0.001Abs (displayed),0.0001Abs(calculated) |
| Light Source | Halogen lamp, 8V/20W |
| Optical filter | 405, 450, 492, 630nm, (4 other filer are optional) |
| Plate Types | Standard with 96-well or other kind of microplate and strip |
| Function | End Point, fixed time, Kinetic |
| Storage | More than 100 programs, up to 100,000 test results |
| Interface | RS-232, USB |
| Dimensions | 454×295×146 mm |
| Weight | 7Kg |
| Electrical Requirement | AC110V/50Hz |
| Catalogue No. | 420250501 |
|
Model |
IMR-4000 |
|
|
Absorbance Range |
0 - 4.000Abs |
|
|
Test Performance |
12 different test in single plate |
|
|
Optical Path |
8-channel |
|
|
Accuracy |
±0.5% or ±0.005Abs |
|
|
Resolution |
0.001Abs (displayed),0.0001Abs(calculated)
|
|
|
Light Source |
Halogen lamp, 8V/20W |
|
|
Band width |
<8nm |
|
|
Optical filter |
405, 450, 492, 630nm, (4 other filer are optional)
|
|
|
Plate Types |
Standard with 96-well or other kind of microplate and strip |
|
|
Function |
End Point, fixed time, Kinetic |
|
|
Reading Speed |
5 seconds for 96 well plate (single wavelength)
|
|
|
Printer |
All printers compatible with Windows
|
|
|
Shaking |
Shaking time and speed adjustable |
|
|
Applications |
ABS, Cut-Off, Single Standard, Curve, Multi-Percent, Percent Log, Linear, Exponential, Power, 4PL Regression |
|
|
Storage |
More than 100 programs, up to 100,000 test results |
|
|
Interface |
RS-232, USB |
|
|
Dimensions |
454×295×146(mm |
|
|
Weight |
7Kg |
|
|
Electrical Requirement |
AC110V/50Hz |
220V/60Hz |
|
Catalogue No. |
420250501 |
420250502 |